生化分析仪系数怎么调 (生化分析仪样本位设置)

生化分析仪读数点设定参考日立 HITACHI 7020

(17s/点)

R1+S R2 R3 测光点: 0-1 5-6 17-18 35 时 间: 0 1.5min 4.5min 10min

日立 HITACHI 7060 /7080

(20s/点)

R1+S R2 R3 测光点: 0-1 4-5 16-17 31 时 间: 0 1.5min 5.5min 10.5min

日立 HITACHI 7170/7180/7600

(20s/点)

R1+S R2 R3 测光点: 0-1 5-6 16-17 34

时 间: 0 1.5min 5.5min 10.5min

奥林帕斯 OLYMPUS AU2700/640/400

(18s/点)

R1+S R2(固定)测光点: 0-1 10-11 27

时 间: 0 3min 8min

奥林帕斯 OLYMPUS AU800/1000

↑ ↑ ↑ (30s/点 )

R1+S R2(固定)测光点: 0-1 7-8 16 时 间: 0 3min 8min

贝克曼 BECKMAN LX

R1 S R2(任选)加 样: -180s 0 ≤620 s 时 间: 0 3min 13min

贝克曼 BECKMAN CX

R1 S R2(任选)加 样: 0 300s 900 s 空白读数: 0(起点)

反应读数: 0(起点) 600 s(≤ 620)时 间: 0 5min 15min

杜邦 DADE Behring AR/RxL

R1 S R2(任选)加 样: -60s 0 600 s 时 间: 0 1min 11min

岛津 Shimadzu CL8000

(17.5s/点 )

R1+S R2(固定)测光点: 0-1 17-18 33 时 间: 0 5min 9.6min

岛津 Shimadzu CL7200/7300

(14s/点)

R1+S R2(固定)测光点: 0-1 16-17 42 时 间: 0 5min 10min

雅培 Abbott Aeroset2000/C8000

( 18.2s/点)

R1+S R2(固定)测光点: 0-1 16-17 33 时 间: 0 5min 10min

欧霸 /蓝宝石 600

(10s/点)

R1+S R2

测光点: 0-1 13 51 时 间: 0 2:10m 8.5min

京都 TMS1024

(13s/点)

R1+S R2(固定)测光点: 0-1 21 46 时 间: 0 5min 10min

注:京都 TMS1024i是它的中文版,读数点有所不同。

东芝 TOSHIBA 40FR

(9.7s/点)

R1 S R2(固定)测光点: 0 7 33:34 62

时 间: 0 1min 5min 10min 注:在 33点前读奇数;在 33点后读偶数。迈瑞 200/300 MINDRAY BS-200/300

-1 0 1 ↑ ↑

R1+S R2 注:迈瑞 200 1min = 4 点 (15s/ 点)

迈瑞 300 1min = 5 点 (12s/ 点)对于酶法起始反应减去终止反应要大于 7点(动力学法 )

迈瑞 400 MINDRAY BS-400

P1 P13 P43 P80 ( 9s/点 )↑ ↑ ↑ 80

R1 S R2 时间 : │← 1min52s →│← 4min30s →│← 5min42s →│

注:终点法 两框最大差读点≤ 5 ; 动力学 两框最小差读点< 3 。

优利特 URIT8000

(22s/点)

R1+S R2 测光点: 0-1 16-17 35

时 间: 0 5-6min 13min

科华卓越 330

(16.7s/点)

R1+S R2

测光点: 0-1 18-19 36 55 时 间: 0 5-6min 10min 15min

科华卓越 400

(18s/点)

R1+S R2 测光点: 0-1 15-16 33 50 时 间: 0 5-6min 10min 15min

西门子 Advia 1200/2400

(14s/点)

R1+S R2 测光点: 0-1 20-22 42

时 间: 0 5-6min 10min

西门子 Advia 1650/1800

(6s/点)

R1+S R2 测光点: 0-1 43-45 98 时 间: 0 5-6min 10min